Titre | Multicenter defocus curve evaluation of a novel trifocal presbyopia correcting IOL – 6 months post-op results |
Abstract Nr. | 2082 |
But | To determine the Binocular Defocus Curve of a novel trifocal lens, the AcrySof IQ PanOptix IOL at 6 months post-op. |
Méthodes | A prospective, single arm, non-randomized, multi-center study was conducted involving the bilateral implantation of AcrySof IQ PanOptix Presbyopia Correcting IOL in 151 subjects. The binocular defocus curves were evaluated under photopic lighting conditions. |
Résultats | A total of 148 subjects (38% M: 62% F) with a mean age of 68.9 ± 9.3 years were implanted bilaterally with the AcrySof IQ PanOptix IOL and had their binocular Defocus Curve assessed at 6-months post-op. The Mean Defocus curve VA from 0.00D to -3.00D ranged from 0.1 to 0.0 logMAR. Average VA at distance (0.00D), intermediate (-1.50D) and near (-2.50D) was 0.0 ± 0.1D, 0.0 ± 0.1D and 0.1 ± 0.1D, respectively. |
Conclusion | The early results from this study show very good visual performance (VA better or equal to 20/25) of the AcrySof IQ PanOptix IOL across the whole range of defocus. |
Conflit d'intérêt | Oui |
Détails conflits d'intérêt | Consultant & study investigator Alcon-Novartis |
Nom | SALLET |
Initiales | G |
Institut | Ooginstituut |
Ville | Aalst |
Nom | KOHNEN |
Initiales | T |
Institut | Goehte University |
Ville | Frankfurt |