Abstract bekijken

Deze abstract is toegekend aan sessie AOB Free papers
TitelMulticenter defocus curve evaluation of a novel trifocal presbyopia correcting IOL – 6 months post-op results
Abstract Nr.2082
DoelTo determine the Binocular Defocus Curve of a novel trifocal lens, the AcrySof IQ PanOptix IOL at 6 months post-op.
MethodesA prospective, single arm, non-randomized, multi-center study was conducted involving the bilateral implantation of AcrySof IQ PanOptix Presbyopia Correcting IOL in 151 subjects. The binocular defocus curves were evaluated under photopic lighting conditions.
ResultatenA total of 148 subjects (38% M: 62% F) with a mean age of 68.9 ± 9.3 years were implanted bilaterally with the AcrySof IQ PanOptix IOL and had their binocular Defocus Curve assessed at 6-months post-op. The Mean Defocus curve VA from 0.00D to -3.00D ranged from 0.1 to 0.0 logMAR. Average VA at distance (0.00D), intermediate (-1.50D) and near (-2.50D) was 0.0 ± 0.1D, 0.0 ± 0.1D and 0.1 ± 0.1D, respectively.
ConclusieThe early results from this study show very good visual performance (VA better or equal to 20/25) of the AcrySof IQ PanOptix IOL across the whole range of defocus.
BelangenverstrengelingJa
Details verstrengelde belangenConsultant & study investigator Alcon-Novartis
Auteur 1
NaamSALLET
InitialenG
InstituutOoginstituut
StadAalst
Auteur 2
NaamKOHNEN
InitialenT
InstituutGoehte University
StadFrankfurt
top ^